Aligning Sensors

Alignment is the process you use to automatically calculate transformations (rotations and translations / offsets) that are applied to a sensor's scan data while it is scanning targets. If you do not correct for these rotations, scan data may be too distorted for your application, and your measurements may therefore be inaccurate. Alignment is often required for various reasons:

With laser line profile sensors, unless you need 6 degrees of freedom, you should use the alignment procedure on the Acquire page, in the Alignment section, by clicking the Align....

Once you click the Align... button, an alignment "wizard" goes through the steps required.

Alignment routine set to Moving Bar alignment, with 3 degrees of freedom selected.

Two ways of performing alignment on sensors are available:

If you are not sure which type of alignment you should you for your application, see Planning Alignment and Choosing an Alignment Method.

Sensors are pre-calibrated and ready to deliver data in engineering units (mm) out of the box. Alignment procedures do not affect sensor calibration.